| 
		 
			MR Safety 
  
		
			
				
				 
  | 
				
				2209.    
				
				  | 
				
				
				
				Magnetically Induced Force Measurements per ASTM F2052 of Active 
				Implantable Medical Device Lead Materials  
					Michael Childers1, Roya Hashemi Rad1, 
					Richard Williamson1, and Shiloh Sison1 
					1St. Jude Medical, Sylmar, CA, United States 
				
					This abstract presents magnetically induced force 
					measurements per ASTM F2052 of materials commonly used in 
					implantable leads.  Implantable leads which are constructed 
					solely from tested materials which pass the magnetically 
					induced force testing acceptance criteria (i.e. gravity 
					force), may not require magnetically induced force testing 
					per ASTM F2052 for MR conditionality with 3 T MR scanners. 
				 
  | 
			 
			
				
				 
				  | 
				
				2210.    
				
				  | 
				
				
				
				Detailing the MR Safety of Intraocular Tantalum Markers Used for 
				Treatment Planning of Proton Beam Therapy of Uveal Melanoma: A 
				7.0T Study  
					Eva Oberacker1, Katharina Paul1, Lukas 
					Winter1, Celal Oezerdem1, Antje Els1, 
					Andreas Pohlmann1, Laura Boehmert1, 
					Stefanie Kox1, Min-Chi Ku1, Till 
					Huelnhagen1, Oliver Stachs2, Jens 
					Heufelder3,4, Andreas Weber3,4, and 
					Thoralf Niendorf1,5 
					1Berlin Ultrahigh Field Facility (B.U.F.F.), Max 
					Delbrück Center for Molecular Medicine in the Helmholtz 
					Association, Berlin, Germany, 2Department 
					of Ophthalmology, University of Rostock, Rostock, Germany,3Department 
					of Ophthalmology, Charité University Medicine, Berlin, 
					Germany, 4BerlinProtonen, 
					Helmholtz Zentrum Berlin, Berlin, Germany, 5Experimental 
					and Clinical Research Center (ECRC), a joint cooperation 
					between the Charité Medical Faculty and the Max Delbrück 
					Center for Molecular Medicine in the Helmholtz Association, 
					Berlin, Germany 
				
					This work examines the MR safety of intraocular tantalum 
					markers used in proton beam therapy of uveal melanoma. RF 
					power deposition induced heating was studied using 
					electromagnetic field and temperature simulations. Magnetic 
					force acting on the marker was investigated and image 
					artifacts were assessed. Minor local increase of RF power 
					deposition was observed for SAR0.075g but 
					not detectable for SAR1g. Measurements showed no 
					detectable magnetic attraction of the implant. FSE based 
					imaging showed only small artifacts barely exceeding the 
					thickness of the sclera. Our studies indicate that 
					intraocular tantalum markers do not constitute a per se 
					contraindication for 7.0T MRI. 
				 
  | 
			 
			
				
				 
				  | 
				
				2211.    
				
				  | 
				
				
				
				SAR/B1+ calibration workflow for safe, high duty-cycle parallel 
				transmission imaging at ultra-high field  
					Filiz Yetisir1, Bastien Guerin2, 
					Lawrence Wald2,3, and Elfar Adalsteinsson1,3 
					1Electrical Engineering and Computer Science, 
					Massachusetts Institute of Technology, Cambridge, MA, United 
					States, 2Dept.of 
					Radiology, Martinos Center for Biomedical Imaging, 
					Charlestown, MA, United States,3Harvard-MIT 
					Division of Health Sciences Technology, Institute of Medical 
					Engineering and Science, Cambridge, MA, United States 
				
					In this work, we propose a pTX safety workflow that will 
					enable high duty cycle imaging at high field systems. 
					Several SAR and B1+ calibration 
					steps are suggested for a complete analysis including 
					modeling the TX array, testing it over time and different 
					loads and finding a safety margin to account for RF system 
					imperfections. Good qualitative agreement was achieved 
					between the simulated and measured B1+ maps 
					for the TX array. 11% and 6° standard deviation was observed 
					in the magnitude and the relative phase maps over time. A 
					maximum difference of 16% was observed between offline and 
					online calculated local SAR values due to RF system 
					imperfections. 
				 
  | 
			 
			
				
				 
  | 
				
				2212.    
				
				  | 
				
				
				
				Direct optical measurement of the RF electrical field for MRI  
					Isabelle Saniour1, Anne-Laure Perrier2, 
					Gwenaël Gaborit2,3, Jean Dahdah3, 
					Lionel Duvillaret3, and Olivier Beuf1 
					1CREATIS, Université de Lyon ; CNRS UMR5220 ; 
					Inserm U1044 ; INSA-Lyon ; Université Claude Bernard Lyon 1, 
					Villeurbanne, France, 2IMEP-LAHC, 
					UMR 5130 ; Université de Savoie, Le Bourget-du-Lac, France,3Kapteos, 
					Sainte-Hélène du Lac, France 
				
					In MRI, a real time monitoring of the magnitude of the 
					electric field prevents the patient from safety hazards due 
					to heating phenomenon. A sub-cm electro-optical probe was 
					used to localize and measure the E-field in 4.7-T MRI. This 
					probe is formed from an electro-optic crystal that changes 
					its refractive indexes according to the applied E-field. The 
					results show that the probe is non-perturbative regarding 
					the E-field and does not affect the quality of MR images. 
					Six clear E-field concentrations were localized at proximal 
					and distal sides of the transceiver coil. Their magnitudes 
					vary between 10000V/m and 20000V/m.    
				 
  | 
			 
			
				
				 
  | 
				
				2213.    
				
				  | 
				
				
				
				On Peripheral Nerve Stimulation of a Compact, Asymmetric 
				Head-Only Gradient Coil: Head Orientation Dependence  
					Seung-Kyun Lee1, Kishore V. Mogatadakala2, 
					Dominic Graziani1, Jean-Baptiste Mathieu2, 
					Thomas K.-F. Foo1, and Matt A. Bernstein3 
					1GE Global Research, Niskayuna, NY, United 
					States, 2GE 
					Healthcare, Florence, SC, United States, 3Mayo 
					Clinic, Rochester, MN, United States 
				
					Head orientation dependence of the peripheral nerve 
					stimulation (PNS) thresholds and the induced electric fields 
					of a high-performance, asymmetric head-only gradient coil 
					were studied experimentally and by numerical simulation. In 
					the experiment, the gradient field direction was fixed and 
					the subject head was rotated in the transverse plane. The 
					subject-reported PNS thresholds nearly doubled when the 
					head's anterior-posterior direction was parallel to the 
					gradient compared to when the head was approximately 
					perpendicular to the gradient. Human-body-model simulation 
					suggested that the orientation dependence may be primarily 
					due to locally concentrated electric fields in the 
					corrugated regions of the face. 
				 
  | 
			 
			
				
				 
  | 
				
				2214.    
				
				  | 
				
				
				
				Positioning to decrease hot spots caused by an intramedullary 
				rod implanted in a forearm  
					Yu Kikuchi1, Minghui Tang1, and Toru 
					Yamamoto2 
					1Graduate School of Health Sciences, Hokkaido 
					university, sapporo, Japan, 2Faculty 
					of Health Sciences, Hokkaido university, sapporo, Japan 
				
					RF heating causes most of incidents during MRI examinations. 
					There still are patients who were implanted metallic 
					products before the advent of MRI and MR compatibility of 
					most such products is unknown.  It was reported that an MRI 
					examination of a patient implanted an intramedullary rod in 
					his forearm was aborted due to a heating claim from the 
					patient. In this study, we confirm RF heating of such 
					patient by using an electromagnetic analysis software 
					dedicated for MRI, and shows that positioning of an 
					implanted arm can decrease SAR sufficiently enough to take 
					MRI examinations. 
				 
  | 
			 
			
				
				 
  | 
				
				2215.    
				
				  | 
				
				
				
				Assessment of Radio Frequency Induced Heating On or Near 
				Implants during MRI – some open issues  
					Mikhail Kozlov1,2 and 
					Gregor Schaefers1,3 
					1MR:comp GmbH, Gelsenkirchen, Germany, 2Max 
					Planck Institute for Human Cognitive and Brain Sciences, 
					Leipzig, Germany, 3Magnetic 
					Resonance Institute for Safety, Technology and Research 
					GmbH, Gelsenkirchen, Germany 
				
					We evaluated locations of maximum temperature rise (max(ΔT)) 
					and the dependence of max(ΔT) 
					on RF-induced power deposition (Ptotal) for 
					some generic implants. ΔT spatial 
					and temporal variations were investigated. To fulfill ASTM 
					F2182-11a setup requirements, the temperature probe should 
					be placed with submillimetre precision at location that 
					cannot be predicted by a full wave electromagnetic 
					simulation alone. It is a challenge to validate with small 
					uncertainty Ptotal calculated 
					using EM simulation by only measuring SAR or VLD value at 
					some points in space, if the field probe sensor size is 
					larger than one tenth of the wire diameter. 
				 
  | 
			 
			
				
				 
  | 
				
				2216.    
				
				  | 
				
				
				
				Reduction of the E field at the tip of implanted wires generated 
				by pTx coils using RF current measurements  
					Gerd Weidemann1, Frank Seifert1, and 
					Bernd Ittermann1 
					1Physikalisch-Technische Bundesanstalt (PTB), 
					Braunschweig and Berlin, Germany 
				
					The possibility to reduce implant heating is an added value 
					option of parallel transmission. An orthogonal-projection 
					method (OPM) is presented to reduce the E fields at the tip 
					of wire type implants by using voltage vectors orthogonal to 
					the vector inducing the worst case RF current at the 
					protruding end of the implant. Experiments confirm that the 
					minimization of RF current at the protruding end leads to a 
					distinct reduction of the electric field at the tip of the 
					wire. Low-hazard steering conditions for n-element 
					pTx coils can be determined in real time during an MR 
					investigation from the measurement of only n complex 
					valued RF currents at the protruding end of the implant. 
				 
  | 
			 
			
				
				 
  | 
				
				2217.    
				
				  | 
				
				
				
				Visualization and Localization of Implanted Devices with 
				Parallel Transmit Array Using Reversed RF Polarization  
					Parnian Zarghamravanbakhsh1, John M Pauly1, 
					and Greig Scott1 
					1Electrical Engineering, Stanford University, 
					Stanford, CA, United States 
				
					The radiofrequency (RF) transmit field can induce current in 
					implanted devices; therefore, it is essential to detect and 
					minimize coupling to stimulator leads and guide-wire 
					structures. Reverse polarization has been proposed as 
					low-RF-power method to safely detect current in the 
					implanted devices using birdcage coil. The purpose of this 
					study is to demonstrate feasibility of combining knowledge 
					of coil current and location with reverse polarization 
					method using parallel transmit array to detect and localize 
					implanted wires.  
				 
  | 
			 
			
				
				 
  | 
				
				2218.    
				
				  | 
				
				
				
				Heating of lead electrodes disconnected from sacral stimulator 
				during routine lumbar MRI at 3T with receive-only coil  
					Pallab K Bhattacharyya1, Howard Goldman2, 
					Mark J Lowe1, Adrienne Quirouet2, and 
					Stephen E Jones1 
					1Imaging Institute, Cleveland Clnic, Cleveland, 
					OH, United States, 2Glickman 
					Urological Institute, Cleveland Clnic, Cleveland, OH, United 
					States 
				
					RF heating testing during lumbar scans of Medtronic 
					Interstim II (Model 3058) implantable pulse generator (IPG) 
					connected to Medtronic Quadipolar Nerve Stimulator Lead 
					(Model 3889) at 3T whole body Siemens TIM Trio scanner with 
					receive-only cervical-lumbar-thoracic coil was performed. 
					Temperatures of the electrodes were measured by using fiber 
					optic sensors with fluoroptic monitoring with the IPG and 
					lead placed inside an ASTM gel phantom. No electrode heating 
					was observed when the lead was connected with the IPG in any 
					of the scans, while considerable heating was observed when 
					the IPG was disconnected and taken out of the phantom. 
				 
  | 
			 
			
				
				 
  | 
				
				2219.    
				
				  | 
				
				
				
				Comparing RF heating simulations and experimental results in pTx 
				coils: an evaluation of three simulation methods  
					Hongbae Jeong1, Peter Jezzard1, and 
					Aaron Hess2 
					1FMRIB Centre, University of Oxford, Oxford, 
					United Kingdom, 2Department 
					of Cardiovascular Medicine, University of Oxford, Oxford, 
					United Kingdom 
				
					In this study, we conducted thermal simulations using EM 
					simulation software and compared these to proton resonance 
					frequency (PRF) thermometry using an ultra-high-field MR 
					phantom. RF heating was measured in the magnet environment 
					using a PRF-based 3D GRE on a 8-channel pTx coil. Three 
					types of simulation method were assessed and compared with 
					experimental data. Amongst the three simulation methods the 
					realistic capacitance simulation was closest to the 
					experimental measurement. In conclusion, PRF RF heating 
					measurements with real fiber optic temperature changes can 
					be used to assess and validate different types of RF 
					simulation.  
				 
  | 
			 
			
				
				 
  | 
				
				2220.    
				
				  | 
				
				
				
				Statistical Equivalence Test Protocol for RF Performance of AIMD 
				Systems  
					Li-Yin Lee1, Shiloh Sison2, Shi Feng3, 
					Kishore Kondabatni4, and Richard Williamson5 
					1BioStatistics, St. Jude Medical, Sylmar, CA, 
					United States, 2Electrical 
					Engineering, St. Jude Medical, Sunnyvale, CA, United States, 3Electrical 
					Engineering, St. Jude Medical, Sylmar, CA, United States, 4St. 
					Jude Medical, Sylmar, CA, United States, 5Program 
					Management, St. Jude Medical, Sylmar, CA, United States 
				
					Test methods for MRI safety and RF safety of AIMD systems 
					has been defined through ISO/TS 10974 are cumbersome to 
					perform on every device and lead combination.  A clear 
					method for determination that two likely equivalent systems 
					has not been described.  The Concordance Correlation 
					Coefficient has been described for this purpose in assay 
					comparison.  This paper evaluates the CCC method for RF 
					equivalence in presence of measurement uncertainty, and 
					confirms that the CCC method is simple and robust for this 
					purpose.     
				 
  | 
			 
			
				
				 
  | 
				
				2221.    
				
				  | 
				
				
				
				Modelling the RF safety of tattoo pigment ink for subjects 
				undergoing 7 Tesla MRI  
					Hongbae Jeong1, Aaron Hess2, and Peter 
					Jezzard1 
					1FMRIB Centre, University of Oxford, Oxford, 
					United Kingdom, 2Department 
					of Cardiovascular Medicine, University of Oxford, Oxford, 
					United Kingdom 
				
					Despite many reports of skin burns in the region of tattoos, 
					there are few safety studies concerning RF heating caused by 
					tattoos. Manufacturers of tattoo ink are numerous and use a 
					range of dye ingredients, making it difficult to assess the 
					electromagnetic properties of each ink pigment. An 
					anchor-shaped tattoo was modelled 1mm under the skin layer 
					in the region of the cervical spine to predict a possible 
					skin burn generated by RF coil. A simulation model of RF 
					heating in tattoo pigment is proposed, which shows that 
					certain tattoo pigments may lead to severe skin burns when 
					performing high field MRI. 
				 
  | 
			 
			
				
				 
  | 
				
				2222.    
				
				  | 
				
				
				
				Heterogeneous gelatin-based head phantom for evaluating DBS 
				heating  
					Clare McElcheran1, Benson Yang2, Fred 
					Tam2, Laleh Golenstani-Rad3, and Simon 
					Graham2 
					1University of Toronto, Toronto, ON, Canada, 2Sunnybrook 
					Health Sciences Centre, Toronto, ON, Canada, 3Massachusetts 
					General Hospital, Charlestown, MA, United States 
				
					A method to create a heterogeneous head phantom with long 
					implanted wires to improve the evaluation of tissue heating 
					surrounding deep brain stimulation (DBS) leads is 
					presented.  The phantom consists of three different 
					oil-in-gelatin dispersions with electrical properties that 
					mimic grey matter, white matter and cerebral spinal fluid 
					(CSF) as well as a human skull.  3D printing technology was 
					used to create gelatin moulds and an acrylic casing.  A CT 
					scan of the human skull was obtained to create a mesh-based 
					digital representation.  Thus, the physical phantom has an 
					associated mesh-based digital model which can be used in 
					electromagnetic simulation. 
				 
  | 
			 
			
				
				 
  | 
				
				2223.    
				
				  | 
				
				
				
				Potentially hazardous materials left behind after an MRI 
				installation  
					Ken E Sakaie1, Wanyong Shin1, and Lowe 
					J Mark1 
					1Imaging Institute, The Cleveland Clinic, 
					Cleveland, OH, United States 
				
					We share our experience discovering and removing metallic 
					objects left behind after a routine MRI hardware upgrade. 
					The results suggest that vigilance is necessary despite the 
					routine nature of such an upgrade. 
				 
  | 
			 
			
				
				 
  | 
				
				2224.    
				
				  | 
				
				
				
				Increased Signal Intensity of brain structures on unenhanced 
				T1-weighted images following 35 or more GBCA administrations  
					Yang Zhang1,2, Yan Cao1, George Shih1, 
					Elizabeth Hecht3, and Martin R Prince1,4 
					1Radiology, Weill Cornell Medical Center, New 
					York, NY, United States, 2Radiology, 
					Qilu Hospital, Shandong University, Jinan, China, People's 
					Republic of, 3Columbia 
					University, New York, NY, United States, 4Radiology, 
					Columbia University, New York, NY, United States 
				
					In 16 patients with 35 or more linear GBCA administrations 
					increased T1 signal on unenhanced images was observed in 
					dentate nucleus (100%), globus pallidus (100%), cerebral 
					peduncles (100%), substantial nigra (88%), red nucleus 
					(88%), colliculi (81%), posterior thalamus (75%), superior 
					cerebellar peduncle (56%), internal capsule (50%), head of 
					caudate nucleus (31%), body of caudate nucleus (25%) , whole 
					thalamus (25%), pons (13%), anterior commissure (13%), 
					posterior brain stem (6%), pituitary gland (6%), mammillary 
					body (6%) and putamen (6%).  The source of T1 signal 
					increase is unknown but may relate to GBCA administration. 
					No clinical significance was identified. 
				 
  | 
			 
			
				
				 
  | 
				
				2225.    
				
				  | 
				
				
				
				Power deposition into a metallic hip prosthesis exposed to 
				switched gradient fields  
					Luca Zilberti1, Oriano Bottauscio1, 
					Mario Chiampi2, Jeffrey Hand3, Hector 
					Sanchez Lopez4, Rüdiger Brühl5, and 
					Stuart Crozier6 
					1Istituto Nazionale di Ricerca Metrologica, 
					Torino, Italy, 2Dipartimento 
					Energia, Politecnico di Torino, Torino, Italy, 3Division 
					of Imaging Sciences and Biomedical Engineering, King’s 
					College London, London, United Kingdom, 4Department 
					of Engineering, Universitas Dian Nuswantoro, Semarang, 
					Indonesia, 5Physikalisch-Technische 
					Bundesanstalt, Berlin, Germany, 6School 
					of Information Technology and Electrical Engineering, 
					University of Queensland, St. Lucia, Australia 
				
					Concern has been recently raised about the possible heating 
					of massive metallic implants, in particular hip prostheses, 
					due to the gradient fields used in MRI. Thus, this 
					contribution discusses the computation of the power density 
					deposited by the magnetic field into the implant, which 
					represents the first step to estimate the thermal heating. 
					The analysis is based on numerical simulations, performed 
					through a computational formulation applied to an anatomical 
					model of the body. The results provide evidence of the role 
					of the three gradient coil axes and of the different 
					harmonic components of the signals in this power deposition 
					process. 
				 
  | 
			 
			
				
				 
  | 
				
				2226.    
				
				  | 
				
				
				
				Testing of a compact ultrasound scanner for use inside clinical 
				interventional MRI suite  
					Chi Ma1, Zaiyang Long1, Diana M 
					Lanners1, Donald J Tradup1, Joel P 
					Felmlee1, David A Woodrum1, Nicholas J 
					Hangiandreou1, and Krzysztof R Gorny1 
					1Department of Radiology, Mayo Clinic, Rochester, 
					MN, United States 
				
					The suitability of a compact Samsung ultrasound (US) system 
					for real-time imaging guidance of treatment device 
					positioning inside 1.5T interventional magnetic resonance 
					imaging (iMRI) suite was assessed. The US system was tested 
					in a proposed site-specific configuration. Magnetic 
					displacement forces exerted by the static magnetic field on 
					each of the US system components were estimated at the 
					proposed locations. Image quality of both MRI and US systems 
					with the US system set to different operating modes were 
					evaluated.  Results demonstrate that this particular US 
					system is suitable for use in the site-specific 
					configuration at our 1.5T iMRI suite.  
				 
  | 
			 
			
				
				 
  | 
				
				2227.    
				
				  | 
				
				
				
				An Evaluation of Radio Frequency Induced Power Deposition of 
				Coaxial Leads with an Implant Model  
					Mikhail Kozlov1,2 and 
					Gregor Schaefers1,3 
					1MR:comp GmbH, Gelsenkirchen, Germany, 2Max 
					Planck Institute for Human Cognitive and Brain Sciences, 
					Leipzig, Germany, 3Magnetic 
					Resonance Institute for Safety, Technology and Research 
					GmbH, Gelsenkirchen, Germany 
				
					We performed 3-D electromagnetic simulations of coaxial 
					leads and numerically obtained the lead models to evaluate 
					power deposition and the voltage induced at the lead 
					proximal end with the lead models. No correlation between 
					peak volume loss density and deposited powers at the tip and 
					the ring was observed. In some cases deposited power at the 
					ring exceeded deposited power at the tip. However further 
					extensive simulations of induced heating behavior should be 
					done before final conclusions regarding coax lead design 
					preferences are made. 
				 
  | 
			 
			
				
				 
  | 
				
				2228.    
				
				  | 
				
				
				
				Influence of electrical properties of lead insulation on radio 
				frequency induced heating during MRI  
					Mikhail Kozlov1,2 and 
					Gregor Schaefers1,3 
					1MR:comp GmbH, Gelsenkirchen, Germany, 2Max 
					Planck Institute for Human Cognitive and Brain Sciences, 
					Leipzig, Germany, 3Magnetic 
					Resonance Institute for Safety, Technology and Research 
					GmbH, Gelsenkirchen, Germany 
				
					We evaluated the dependence of RF-induced power deposited at 
					a hot spot (p) on insulating 
					electrical properties for insulated stainless steel wires of 
					1.5 mm in diameter with insulation thickness of 0.5 mm. Lead 
					transfer functions (TF) were obtained by 3-D electromagnetic 
					simulations. TF and p depended 
					significantly on electrical properties of insulation. 
					Increased insulator conductivity resulted in decreased p. 
					For all insulated wires investigated non-uniform RF 
					excitation resulted in higher power deposition than uniform 
					RF excitation. 
				 
  | 
			 
			
				
				 
  | 
				
				2229.    
				
				  | 
				
				
				
				Design and simulation of a nested 4 channel 1H and 3 channel 13C 
				coil for glycogen NMR experiments in the calf muscle at 7 T  
					Sigrun Goluch1,2,3, Roberta Kriegl2,3, 
					Elmar Laistler2,3, Martin Gajdošík 4,5, 
					and Martin Krššák 1,4,5 
					1Division of Endocrinology and Metabolism, 
					Department of Internal Medicine III, Medical University of 
					Vienna, Vienna, Austria, 2MR 
					Center of Excellence, Medical University of Vienna, Vienna, 
					Austria, 3Center 
					for Medical Physics and Biomedical Engineering, Medical 
					University of Vienna, Vienna, Austria, 4High-Field 
					MR Center, Department of Biomedical Imaging and Image-Guided 
					Therapy, Medical University of Vienna, Vienna, Austria, 5Christian 
					Doppler Laboratory for Clinical Molecular MR Imaging, 
					Medical University of Vienna, Vienna, Austria 
				
					Due to the inherently low sensitivity of carbon-13 NMR, 13C 
					spectroscopic experiments at 7T require specifically 
					optimized double tuned local RF transceive arrays for high 
					SNR, exhibiting sufficient electrical isolation between the 
					arrays to enable 1H decoupling and high SAR efficiency as to 
					not invoke SAR limits during proton decoupling. In this work 
					we present the simulation and optimization of a 7 channel 
					nested 1H/13C 
					RF transceive coil array for 13C 
					metabolic studies in the human calf muscle at 7 T. 
				 
  | 
			 
			
				
				 
  | 
				
				2230.    
				
				  | 
				
				
				
				Assessment of RF induced heating of intracranial Micro-depth 
				electrodes during MRI  
					Anastasia Papadaki1,2, David Carmichael3, 
					Andrew McEvoy4,5, Anna Miserocchi4,5, 
					Tarek Yousry1,2, Beate Diehl4,6, Louis 
					Lemieux4, and John S Thornton1,2 
					1Lysholm Department of Neuroradiology, National 
					Hospital for Neurology and Neurosurgery, UCLH, London, 
					United Kingdom, 2Department 
					of Brain Repair and Rehabilitation, UCL Institute of 
					Neurology, London, United Kingdom, 3Imaging 
					and Biophysics Unit, UCL Institute of Child Health, London, 
					United Kingdom, 4Department 
					of Clinical and Experimental Epilepsy, UCL Institute of 
					Neurology, London, United Kingdom,5Department of 
					Neurosurgery, National Hospital for Neurology and 
					Neurosurgery, London, United Kingdom, 6Department 
					of Neurophysiology, National Hospital for Neurology and 
					Neurosurgery, London, United Kingdom 
				
					In this study we assessed temperature changes (?T) during 
					MRI in the vicinity of microwires EEG electrodes in a 
					phantom. Measurements were performed at 1.5T during a high 
					SAR TSE sequence for two different depth electrode 
					arrangements with and without microwires. Although we 
					observed a small temperature rise due to the presence of 
					microwires the maximum temperature change ?T did not exceed 
					1°C at 1.5T.  
				 
  | 
			 
			
				
				 
  | 
				
				2231.    
				
				  | 
				
				
				
				SAR and patient orientation for 3 T 2-channel parallel transmit 
				pelvis imaging  
					Mariya Lazebnik1 
					1GE Healthcare, Waukesha, WI, United States 
				
					This work investigates the impact of patient orientation on 
					SAR for 3 Tesla two-channel parallel transmit (pTx) pelvis 
					imaging. SAR simulations were performed on two human body 
					models in a supine position in a 70 cm-diameter 3 T body 
					coil in a pelvis landmark, in both “head first” and “feet 
					first” patient entry orientations. Whole body SAR, peak 
					spatial SAR, and SAR ratio (= peak SAR / whole body SAR) 
					were computed for quadrature and pTx excitations. Patient 
					position and orientation can cause peak SAR and SAR ratio to 
					vary significantly and must be considered when evaluating 
					pTx excitation. 
				 
  | 
			 
			
				
				 
  | 
				
				2232.    
				
				  | 
				
				
				
				Subject-specific SAR prediction in adults and children at 7.0T  
					Gianluigi Tiberi1,2, Mauro Costagli1,2, 
					Laura Biagi2, Alessio De Ciantis3, 
					Nunzia Fontana4, Riccardo Stara5,6, 
					Mark R Symms7, Mirco Cosottini8, Renzo 
					Guerrini3, and Michela Tosetti1,2 
					1Imago7, Pisa, Italy, 2IRCCS 
					Stella Maris Foundation, Pisa, Italy, 3Meyer 
					Children’s Hospital, Firenze, Italy, 4Dipartimento 
					di Ingegneria dell'Informazione, Pisa, Italy, 5National 
					Institute of Nuclear Physics (INFN), Pisa, Italy, 6Stanford 
					University, Stanford, CA, United States, 7General 
					Electric ASL Scientist (EMEA), Pisa, Italy, 8Department 
					of Translational Research and New Surgical and Medical 
					Technologies, Pisa, Italy 
				
					In this study we propose a procedure which allows the 
					prediction of global and local subject-specific SAR exposure 
					for commonly used 7.0T sequences. Prerequisites for such 
					prediction are: sequences’ SAR exposure simulated on the 
					generic anatomic models; subject-specific measured B1+ maps. 
					Validation has been provided through phantom experiment. We 
					observed that: SILENT and FLAIR can be safely used in all 
					subjects, both adults and children; FLAIR is more SAR 
					demanding than SILENT; predicted SAR exposure does not show 
					a significant variation with subject weight. 
				 
  | 
			 
			
				
				 
  | 
				
				2233.    
				
				  | 
				
				
				
				Safety of MR Imaging of Patients with Cardiac Implanted Devices  
					El-Sayed H. Ibrahim1, Laura Horwood1, 
					Jadranka Stojanovska1, Luba Frank1, 
					Anil Attili1, Hakan Oral1, and Frank 
					Bogun1 
					1University of Michigan, Ann Arbor, MI, United 
					States 
				
					This study examines whether MRI is safe in patients with 
					cardiac implantable electronic device (CIED) excluded from 
					published protocols, e.g. patients with abandoned leads or 
					pacemaker dependency. A total of 162 MRI scans were obtained 
					in 142 consecutive patients with CIED’s. Cardiac scans were 
					performed in 94 patients and spinal/brain scans were 
					performed in 47 patients. Only one patient developed 
					ventricular tachycardia during a spine scan and was removed 
					from the scanner for device reactivation without 
					consequences. No other adverse events were noted. The 
					devices interrogated parameters essentially remained the 
					same immediately, 1-week after, and 3-months after the 
					scans.  
				 
  | 
			 
			
				
				 
  | 
				
				2234.    
				
				  | 
				
				
				
				The Potential for Eddy Current Induced Peripheral Nerve 
				Stimulation from an Active Implanted Device Canister  
					Xin Chen1, Jonathan Edmonson2, and 
					Michael Steckner1 
					1Toshiba Medical Research Institute USA, Inc., 
					Mayfield Village, OH, United States, 2Medtronic 
					CRHF, Mounds View, MN, United States 
				
					We used numerical simulations to investigate the potential 
					for increased PNS likelihood with implanted device. Modeling 
					of a gradient coil loaded with a human subject with a 
					metallic implanted canister showed that the electric field 
					around the device can increase by up to 3 fold, suggesting 
					increased PNS likelihood. 
				 
  | 
			 
			
				
				 
  | 
				
				2235.    
  | 
				
				
				
				Prospective Observational Post-marketing Study on the Safety of 
				Gadoterate Meglumine - Final Results in the pediatric cohort of 
				over 1,600 children  
					Yun Peng1 
					1Beijing Children's Hospital, Beijing, China, 
					People's Republic of 
				
					An observational post-marketing study was conducted in 10 
					countries to prospectively collect safety data in adults and 
					children who were scheduled to undergo routine Magnetic 
					Resonance Imaging (MRI) with administration of gadoterate 
					meglumine (Dotarem®). The incidence of Nephrogenic Systemic 
					Fibrosis (NSF) in routine practice was assessed through 
					specific follow-up of patients with moderate to severe renal 
					impairment. Final results in a large pediatric 
					sub-population of over 1,600 children showed a very good 
					safety profile of gadoterate meglumine with only one adverse 
					event reported in a child and no suspicion of NSF reported. 
				 
  | 
			 
			
				
				 
  | 
				
				2236.    
				
				  | 
				
				
				
				Scanner-specific verification of Transmit RF Body Coil B1-field 
				to inform clinical triage of patients with implanted devices  
					Chi Ma1, Krzysztof R Gorny1, 
					Christopher P Favazza1, Robert E Watson1, 
					and Heidi A Edmonson1 
					1Radiology, Mayo Clinic, Rochester, MN, United 
					States 
				
					Exclusion of scanning with transmit RF body coil may 
					prohibit access to life-saving diagnoses for patients with 
					MR-conditional implantable devices. Manufacturer provided 
					plots of RF B1-field indicate that RF energy over the 
					implant may be significantly reduced if the implant is kept 
					outside of the 50-55cm long transmit RF body coil.  
					Scanner-specific B1-field measurements and RF-induced 
					heating measurements confirm reduction in heating as 
					conductive material moves away from scanner isocenter.  
					B1-field measurements lateral to the central scanner axis 
					demonstrate local peaks in the B1-field that would not be 
					identified from the IEC-required manufacturer plots. 
				 
  | 
			 
			
				
				 
  | 
				
				2237.    
				
				  | 
				
				
				
				Magnetic Displacement Force and Safety of Coronary Artery Stents 
				at 7 Tesla.  
					Christian Hamilton-Craig1,2, Jess Cameron1, 
					Gregory Brown1, and Graham Galloway1,3 
					1Centre for Advanced Imaging, University of 
					Queensland, Brisbane, Australia, 2Richard 
					Slaughter Centre of Excellence in CVMRI, The Prince Charles 
					Hospital, Brisbane, Australia, 3Translational 
					Research Institute, Brisbane, Australia 
				
					Currently, there are minimal data regarding the magnetically 
					induced displacement force of coronary artery stents, in 7.0 
					T MR.  We tested a range of commonly implanted coronary 
					artery stents for maximal displacement force at 7T.  CoCr 
					stents appear to have safe deflection properties at 7T. 
					However 316L-SS and PtCr stents exhibit increased 
					magnetically induced displacement forces, and may be not be 
					considered conditionally safe at 7.0T 
				 
  | 
			 
			
				
				 
  | 
				
				2238.    
				
				  | 
				
				
				
				Electrocorticography grids might cause excessive heating during 
				MR imaging  
					Emad Ahmadi1, Reza Atefi1, Emad 
					Eskandar2, Alexandra J. Golby3, 
					Michael H. Lev1, Rajiv Gupta1, and 
					Giorgio Bonmassar1 
					1Radiology, Massachusetts General Hospital, 
					Boston, MA, United States, 2Neurosurgery, 
					Massachusetts General Hospital, Boston, MA, United States, 3Neurosurgery, 
					Brigham and Women's Hospital, Boston, MA, United States 
				
					Electrocorticography grids are routinely implanted over the 
					cortex for pre-surgical planning in epilepsy surgery. We 
					propose that MR imaging at 3T might cause heating injury in 
					patients  with implanted electrocorticography grids. 
				 
  | 
			 
			
				
				 
  | 
				
				2239.    
				
				  | 
				
				
				
				Sugar free tissue-mimicking MRI phantoms for improved 
				signal-to-noise ratio  
					Carlotta Ianniello1,2, Ryan Brown1, 
					Martijn Cloos3, Qi Duan4, Jerzy 
					Walczyk3, Graham Wiggins3, Daniel K 
					Sodickson2,3, and Riccardo Lattanzi2,3 
					1Radiology, Center for Advanced Imaging 
					Innovation and Research (CAI2R) and Center for Biomedical 
					Imaging, Department of Radiology, New York University School 
					of Medicine, New York, NY, United States, 2The 
					Sackler Institute of Graduate Biomedical Science, New York 
					University School of Medicine, New York, NY, United States, 3Center 
					for Advanced Imaging Innovation and Research (CAI2R) and 
					Center for Biomedical Imaging, Department of Radiology, New 
					York University School of Medicine, New York, NY, United 
					States, 4Laboratory 
					of Functional and Molecular Imaging, NINDS, National 
					Institutes of Health, Bethesda, MD, United States 
				
					We investigated Polyvinylpyrrolidone (PVP) as an alternative 
					to sugar to control relative permittivity in 
					tissue-mimicking MR phantoms. We constructed a 
					two-compartment phantom filled with water solutions of PVP 
					and NaCl, the latter used to control conductivity. A lower 
					amount of PVP than sugar is required, allowing low 
					permittivity materials to be realized. While signal 
					decreases rapidly in sugar-based phantoms, PVP materials 
					have long T2*/T2, making 
					PVP-based phantoms suitable for the validation of MR-based 
					electrical properties mapping techniques that rely on high 
					SNR of signal and B1+ maps. 
					PVP solutions are relatively inexpensive, easy to mix and do 
					not require preservatives.  
				 
  | 
			 
			
				
				 
  | 
				
				2240.    
				
				  | 
				
				
				
				Experimental evaluation of heating and SAR reduction with a 
				dielectric insert at 3T  
					Christopher Sica1, Sebastian Rupprecht1, 
					and Qing X Yang1 
					1Radiology, Penn State College of Medicine, 
					Hershey, PA, United States 
				
					Prior work has suggested that a dielectric insert can reduce 
					the SAR in the brain at 3T. These previous results were 
					obtained via electromagnetic simulations. Here, we present 
					an experimental evaluation of SAR reduction in a phantom 
					with a dielectric insert. 
				 | 
			 
		 
		 |