Robust method for Whole Body DWIBS applied both Image based B0 Shimming and Blip-up Blip-down Distortion Correction
Hiroshi Hamano1, Masami Yoneyama1, Yasutomo Katsumata2, Kazuhiro Katahira3, and Kenji Iinuma1
1Philips Japan, Tokyo, Japan, 2Philips Healthcare, Tokyo, Japan, 3Department of Radiology, Kumamoto Chuo Hospital, Kumamoto, Japan
DWIBS sometimes suffers from sever image distortion due to the presence of air within and/or at edge of the FOV. We demonstrated that whole body DWIBS applied both image based B0 shimming and blip-up blip-down distortion correction to provide higher robustness.
Figure 2. The MIP of direct coronal whole body DWIBS at 3.0T were shown. Image based B0 shimming provided improving depiction of axillary lymph nodes (red circle). Blip-up blip down distortion correction also decrease distortion effect due to the presence of air (blue circle).
Figure 4. The MIP of transverse DWIBS at 1.5T were shown. Similar findings to 3.0T were obtained at 1.5T.