1790
Simultaneous acquisition of 1H/19F/23Na/31P MR imaging in Phantom at 3T using a quadruple-nuclear RF coil system
Nan Li1,2, Xing Yang1,2, Feng Du1,2, Kang Yan3, Bei Liu3, Yiping Du3, Chunsheng Yang4,5, Zhi Zhang4,5, Li Chen4,5, Fang Chen4,5, Xiaoliang Zhang6, Xin Liu1,2, Hairong Zheng1,2, and Ye Li1,2
1Lauterbur Imaging Research Center, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China;, shenzhen, China, 2Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province, Shenzhen, China, Shenzhen, China, 3Institute for Medical Imaging Technology, School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China., Shanghai, China, 4State Key Laboratory of Magnetic Resonance and Atomic and Molecular Physics, Wuhan Center for Magnetic Resonance, Innovation Academy for Precision Measurement Science and Technology, the Chinese Academy of Sciences, Wuhan 430071, China;, Wuhan, China, 5University of Chinese Academy of Science, Beijing 100049, China, Wuhan, China, 6Department of Biomedical Engineering, State University of New York at Buffalo, NY, United States, Buffalo, NY, United States
 Imaging tests were performed to demonstrate the feasibility of synchronized 1H / 19F / 23Na/ 31P MR imaging at 3T by using developed quadruple-nuclear RF coil. The imaging approach combined with triple-tuned 19F / 23Na/ 31P receive coil also indicated the improved SNR.
Fig. 3. (a)-(b): The 1H/19F/23Na/31P MR measured images obtained with the specific phantom. (a) only scanned by using the quadruple-nuclear RF coil. (b) Combined the quadruple-nuclear coil and triple tuned local 19F/23Na/31P receive coil. (c)-(d): SNR maps for the 1H/19F/23Na/31P imaging. (c) Without the local triple-tuned RF coils. (d) Combined with the local triple-tuned 19F/23Na/31P receive RF coils. The position of the local receive RF coil was indicated by the white dotted-line, and the SNR of the region indicated by the red arrow is significantly improved.
Fig. 4. (a)-(b): The 1H/19F/23Na/31P MR measured images obtained with the independent nuclide test phantom. (a) only scanned by using the quadruple-nuclear RF coil. (b) Combined the quadruple-nuclear coil and triple tuned local 19F/23Na/31P receive coil. (c)-(d): SNR maps for the 1H/19F/23Na/31P imaging. (c) Without the local triple-tuned RF coils. (d) Combined with the local triple-tuned 19F/23Na/31P receive RF coils. The position of the local receive RF coil was indicated by the white dotted-line, and the SNR of the region indicated by the red arrow is significantly improved.