2281
RF Impedance of MR-Conditional Pacemaker Leads when Connected to Implantable Pulse Generators from Different MR-Conditional Systems
David Prutchi1, Jason Meyers1, and Ramez Shehada2
1Impulse Dynamics (USA) Inc., Marlton, NJ, United States, 2Medical Technology Laboratories, La Mirada, CA, United States
The IPG’s contribution to the limitation of RF currents is relatively small (1.03 to 8.04%, mean 2.72%).  A maximum difference of 4.85% was found for a hybrid system, which is probably negligible when considering lead impedance variability.
Figure 1: Contribution of IPG input impedance against the total lead/IPG impedance to 63.87 MHz RF currents at the lead’s Tip electrode. Lead/IPG pairs are MR-conditional labeled for whole-body MRI at 1.5T.
Table 1: Contribution of IPG input impedance against the total lead/IPG impedance to 63.87 MHz RF currents at the lead’s Tip electrode. Lead impedance incorporates the effect of the implantable device’s enclosure in gel slurry. Lead/IPG pairs are MR-conditional labeled for whole-body MRI at 1.5T.