Analysis of common-mode rejection ability and Insertion Loss of Dual band Lattice Balun with non-ideal components
Yue Zhu1,2 and Xinqiang Yan1,2
1Vanderbilt University Institute of Imaging Science, Nashville, TN, United States, 2Department of Radiology and Radiological Sciences, Vanderbilt University, Nashville, TN, United States
A dual band Lattice balun was proposed to remove common mode current for both frequencies in dual-tuned MRI coils with a single interfacing unit. In this work, we analyzed how each lumped element will affect this device's performance and optimized the circuit performance based on the analysis. 
Figure 1. Diagram of the dual tuned balun and approximations for different nuclei. The dual-band balun approximates two different single-band baluns at different frequencies, making it suitable for dual-band detection systems.
Figure 5. Common mode rejection ratio test with current probes. The CMRR shows a good rejection ratio at sodium and proton frequencies. The two current probes are connected to the VNA while the ground of the two cables from the balun circuit connects to a large resistor.