2510
Increased B1+ efficiency of a dipole antenna compared to a loop coil for 31P-MRS at 7T: simulations and cardiac MRSI data
Jabrane Karkouri1, Saba Shirvani1, Tiger Zhang1, Dennis Klomp2, Martijn Lunenburg3, Ladislav Valkovic4, and Christopher T. Rodgers1
1Wolfson Brain Imaging Center, University of Cambridge, Cambridge, United Kingdom, 2Department of Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 3Tesla Dynamic Coils, Zaltbommel, Netherlands, 4Oxford Centre for Clinical Magnetic Resonance Research, University of Oxford, Oxford, United Kingdom
We evaluated the performance of two RF coil configurations for ultra-high field 31P-MRS applications in the body via simulations, through phantom and in vivo scans. Both verifications indicate that there is improvement in terms of B1+ efficiency for the dipole + loop vs quadrature loops.
Figure 3: B1+ field simulation results of the dipole coil over a homogeneous phantom and a heart numerical model.
Figure 5: Left: cardiac mid-short axis localiser and CSI grid. Right: mid-septal 31P cardiac spectrum (apodised by 30 Hz) from the voxel highlighted in the CSI grid.