S12
The Effect of Enlarging Shim Volume on B0 homogeneity of MRS Voxel on 7T
Huijun Liao1, Eduardo Jorge Uribe Coello1, Wufan Zhao1, Han Sam Jiang1, and Alexander Lin1
1Center for Clinical Spectroscopy, Brigham and Women's Hospital, Boston, MA, United States
Enlarging shim volume by more than two folds in inhomogeneous regions like left hippocampus and left thalamus did not improve the B0 homogeneity of the MRS voxel. NAA and Glu were not as sensitive to B0 homogeneity as the low concentration metabolites like GSH and GABA.
Figure 1: Correlation between Shim Volume size and FWHM (Hz) reported by Siemens Syngo (in left temporal lobe region).